Section 4: Device Description
▪ The frame selects parameters that can be changed.
▪ This indicator shows the skin Cut-Off temperature, which is
adjustable between 35-42
o
C, with increments of 1
o
C in the
tissue. The selected value indicates measured temperature
which is maintained during the treatment according to the
setting.
▪ This indicator shows the internal Cut-Off temperature, which
is adjustable between 50-70
o
C, with increments of
1
o
C in the tissue. The selected value indicates measured
temperature which is maintained during the treatment
according to the setting.
External Temp.
Internal Temp.
▪ This indicator shows measured external temperature.
▪ This indicator shows measured internal temperature.
▪ This indicates the time that Cut-Off temperature is
maintained. It varies from 15-120 sec with increments of 5
sec.
▪ This indicator shows elapsed treatment time.
▪ The System has three treatment modes: Standby, Ready and
Active.
▪ Standby mode allows the user to set treatment parameters.
Activation of energy is unavailable in Standby mode.
▪ In Ready mode, the system waits for a signal from the
footswitch to activate the energy. Any change to the
treatment settings switches the system to Standby mode.
▪ Active mode is entered during RF energy delivery.