Inspire System Models 3024, 4063, 4323 English 29
200-079-101 Rev A
3024EN_ch.fm 5/6/14 10:31 pm
4.625 x 6 inches (117 mm x 152 mm)
Inspire Medical Confidential
Connecting the Leads and IPG
Connect the respiratory sensing lead to the IPG
1. Wipe off any body fluids from the respiratory sensing lead connector.
2. Grasping the lead approximately 3 cm (1.2 in) from the connector end, insert the lead
connector into the IPG connector port marked SENSE (Figure 15).
• Make sure the lead connector is fully inserted into the IPG connector port by verifying
that the lead connector pin reaches the back of the IPG connector port cavity.
Figure 15. Insert the respiratory sensing lead connector into IPG connector port marked
SENSE
3. Use the white-handled hex wrench to tighten the 2 setscrews adjacent to the SENSE port
(Figure 16) until resistance is felt, then tighten each screw 1/4 turn more. Do not over
tighten.
• Tighten the setscrew furthest from the lead port first.
• Prior to tightening the setscrew nearest to the lead port, gently tug the lead to confirm
that the first setscrew has secured the lead in place.
• After tightening both setscrews, confirm that the seals covering the setscrews are fully
closed.
Caution: Saline or bodily fluids in the IPG connector may reduce battery longevity.
• Do not allow saline or bodily fluids to enter the IPG connector ports.
• Confirm that lead connectors are dry prior to inserting them into the IPG ports.
• Use care when inserting the hex wrench to avoid damage to the seals.
• Confirm that setscrew seals fully close after securing the lead in place.