Inspire System Models 3024, 4063, 4323 English 43
200-079-101 Rev A
3024EN_ch.fm 5/6/14 10:31 pm
4.625 x 6 inches (117 mm x 152 mm)
Inspire Medical Confidential
Configurable Settings
The parameters in Table 10 can be changed using an Inspire programmer. See the physician
programmer manual for more information.
Table 10. Inspire II IPG (Model 3024)
Configurable Settings
Parameter Values Increment
Stimulation
Start Delay 0–75 mins 5 mins
Pause Time 5–30 mins 5 mins
Therapy Duration 1–15 hrs 1 hr
Amplitude 0.0–5.0 V 0.1 V
Rate 20, 25, 30, 33, 40 Hz
Pulse Width 60, 90, 120, 150, 180, 210 μs
Amplitude Ramp 0.0–0.5 sec 0.125 sec
Electrode Configuration Outer (+) Center (-)
Outer (-) Center (+)
Case (+) Outer (-)
Case (+) Center (-)
Case (+) Outer (-) Center (-)
Patient Amplitude Control On, Off
Sensing
Exhalation Sensitivity -4 to +3 1
Exhalation Threshold -1, 0, +1 1
Inhalation Sensitivity -7 to +1 1
Inhalation Threshold 0, +1 1
Hard Off Period 38, 50, 63, 75%
Soft Off Period 13, 25%
Invert Signal On, Off
Max Stim Time 2–4 secs 1.0 sec