Loading...
Inspire Sleep Remote English 45
You do not hear
audio tones, but
the sleep remote
is working.
Audio tones are
turned off.
Turn audio tones on (refer to
page 19).
You hear three
beeps after
pressing the
Increase ( ) or
Decrease button
().
The stimulation
strength is at the
upper or lower
limit.
Confirm that the current
stimulation strength does not
work for you. Contact your
doctor to change the
stimulation strength limits.
You hear three
beeps several
seconds after
pressing a
button.
A setting in your
generator has
changed.
Communication
with your
generator was
not successful.
Review current therapy
settings and try your change
again.
Stimulation is
uncomfortable.
1. Press the Therapy Off button
().
2. Reduce the stimulation
strength by pressing the
Decrease button ( ), and
then holding the sleep
remote over your generator.
3. Press the Therapy On button
( ) to test the stimulation
strength. Decrease
stimulation further if
necessary (refer to page 29).
If this happens: Possible
causes:
Try these steps: