EasyManua.ls Logo

Instek Digital LCR-821 - Maintenance and Calibration Recall; Recall Calibration Function Steps

Default Icon
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LCR-800 Series User Manual
46
7-3. Recall Calibration
When the measured value is inaccurate, you can use the Recall Calibration
function to rework by proceeding the steps as follows:
z Press MENU key. (Figure 7-1)
z Press F3 key to select “SET PARAMETER” function.Figure 7-2
z Press F4 key to select “RECALL CALIBRATION” function.Figure 7-3
z Press compound Key 1 to select the “Yes” to start recall calibration.
Figure 7-4
z Press compound Key 2 to select the “No” to stop recall calibration.
(Figure 7-4)
If the function keys are not active, do not power off and wait another 2
minutes, then try again.
Figure 7-1
SPEED
SLOW
DISPLAY
VALUE
MODE
C/D
CIRCUIT
SERIES
MENU
C 13.450uF
D .1652
TESTING
F :1.0000 kHz
V :1.000 V
AUTO MANU INT.B OFF
R.H OFF
C.V OFF
Press menu
key
LCR-800 Series User Manual
47
OFFSET
SORT
SETTING
CALBRAT
EXIT
CAP. C/D OFFSET
SET SORT
SET PARAMETER
CALIBRATION
Press F3 key
to the setting
menu
VOLT
77.000
EXIT
MEMMORY NO: 0
VOLTAGE= 1.000 V
AVERAGE= 1
RECALL CALIBRATION
Press F4 key to
input the recall
function
RECALL
NOM.VAL
1
MEMORY
0
1.000
VOLT
77.000
EXIT
MEMMORY NO: 0
VOLTAGE= 1.000 V
AVERAGE= 1
RECALL CALIBRATION
Are you sure?(1)Yes
(2)No
RECALL
NOM.VAL
1
MEMORY
0
1.000
MEMORY
Figure 7-2
Figure 7-3
Figure 7-4

Related product manuals