7. Troubleshooting
CoreDry does not turn on –
Make sure unit is plugged in then test by plugging
another device into the same outlet such as a lamp or
fan.
Power supplies may not be working.
Contact InstroTek for troubleshooting help 919-875-8371.
CoreDry runs but does not draw vacuum –
Make sure both lids are seated correctly on the sample
tank and the vapor/cold trap. Push down gently to seat
while unit is running.
Install clean tank port fabric filters.
Make sure the pump oil fill cap is installed, fitting
correctly and tight.
Check oil drain valve, should be closed completely.
Run Self-Test to make sure everything is working within
normal parameters.
Open CoreDry and make sure no hoses are pinched or
loose.
Contact InstroTek for troubleshooting help 919-875-8371.
CoreDry works but is taking longer time to dry samples than
before –
Make sure lids, pump fill cap and drain valve are all
correctly sealed.
Install clean tank port fabric filters.
Check the vacuum readings on the screen for a dry
empty chamber. The readings should be 6mmHg or
lower.
Run Self-Test to make sure everything is working within
normal parameters.
Feel vapor trap to make it is sure cold.
Change the pump oil. Test to see if results have
improved.
Change pump exhaust filter. Test to see if results have
improved.
Contact InstroTek for troubleshooting help 919-875-8371.