Table 204: Guidance and Manufacturer’s Declaration – Emissions Equipment and Systems
that are NOT Life-Supporting
Table 206: Recommended Separation Distances Between Portable and Mobile RF Communi-
cations Equipment and the MLX Equipment and Systems that are NOT Life-Supporting
Electromagnetic Compatibility (EMC) User Information
The MLX light source is intended for use in the electromagnetic environment specified below. The customer or user of the MLX
light source should assure that it is used in such an environment.
The MLX light source is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The
customer or user of the MLX light source can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF Communications Equipment and the MLX light source as recommended below, according to
the maximum output power of the communications equipment.
Immunity Test
IEC 60601 Test Level Compliance Level
Electromagnetic Environment - Guidance
Conducted RF
IEC 61000-4-6
3 Vrms from
150 kHz to 80 MHz
V1 = 3 Vrms Portable and mobile RF communications equipment should
be separated from the MLX light source by no less than the
recommended separation distances calculated/listed below:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
E1 = 3V/m
80 to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum power rating in was and D is the
recommended separation distance in meters.
Field strengths from fixed transmiers, as determined by an
electromagnetic site survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the vicinity of equipment
containing a transmier
Recommended Separation Distances for the MLX (meters)
Maximum Output Power
(Was)
150 kHz to 80 MHz 80 to 800 MHz
800 MHz to 2.5 GHz
0.01 0.11667 0.11667 0.23333
0.1 0.36894 0.36894 0.73785
1 1.1667 1.1667 2.3333
10 3.6894 3.6894 7.3785
100 11.667 11.667 23.333
16