5. The system will start up at the same light intensity as when last used. For new systems, the light intensity will start at the
minimum 20% seing.
6. To adjust the light intensity, push the Membrane Switches (Figure 4): + to increase or - to decrease the light output or
intensity. Range is 0%, 20 – 100% in 5% increments. Holding the + or – buon will change the intensity more quickly.
7. The System Status Display can be pressed to show both lamp time and system operation time. A second press of the switch
will show the soware version of the system monitor. A third press returns to light output. (The system will automatically
return to the light intensity reading aer 15 seconds).
8. The system may be placed in and out of stand-by mode by using the stand-by switch. The system will emit no light in this
state and allow the surgeon to unplug a headlight, move around the table, or change cables in a laparoscopic or
endoscopic procedure.
9. Take precautions not to touch or disconnect the cable end fiing
from the turret until the light source has been “shut down” for a period
of time and allowed to cool. The fiber optic cable end fiing will remain
hot immediately following shut down, which can cause burns.
10. To shut the system off, simply press the main Power Switch as in Step 2.
Note
It is strongly recommended that the light be used at the
minimum intensity for good visualization.
Note
If unit is le in stand-by mode for 15 minutes, it
will automatically shut off the lamp. Fans will keep
running. To re-activate, push the Stand-by Buon.
Caution
Fiber Optic Cable end fiing can be hot
7
Integra
®
Luxtec
®
MLX 300 Wa Xenon Light Source
membrane switches
system status display
4