INTELECT LEGEND STIM USER MANUAL
22
Stimulator Output Parameters
Parameter Russian High Volt
Function Mode Electrodes Electrodes, Probes
Single, Recipr. Single
Co-Contraction
Carrier Frequency 2500 Hz N/A
Pulse Frequency N/A 10-120 pps
Burst Frequency 20-100 BPS N/A
Phase Duration N/A N/A
Interphase Interval N/A N/A
Duty Cycle 10-50% N/A
Ramp Up / Ramp Down .5, 1, 2, 5 sec N/A
Cycle Time 5/5,10/10, 10/20 5/5,10/10, 10/20
4/12, 10/30, 0/50 4/12, 10/30, 0/50
Continuous Continuous
Polarity N/A Pos. (+), Neg. (-)
Amplitude 0-100 mA RMS 0-500 mA RMS
into 500 ohm load
Voltage (max) 200 Volts 0-500 Volts
Output Current N/A
0-2500 mA Peak
Treatment Time 0-60 min 0-60 min
N/A = Not Applicable