EasyManua.ls Logo

Intelect NMES - 5.9 Adjust Relaxation (OFF) Time; 5.10 Adjust the Pulse Rate; 5.11 Adjust the Pulse Width; 5.12 Adjust Timer

Intelect NMES
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.9 Adjust Relaxation (OFF) Time:(except the Constant Mode)
From 2 seconds to 99 seconds, press SET button to enter the OFF Time setting mode, then press INCREASE or
DECREASE button to adjust OFF Time to the setting recommended by your medical professional. In order to
prevent the spasticity, the Relaxation Time cannot be adjusted less than the Contraction Time.
5.10 Adjust the Pulse Rate:
The pulses rate are adjustable 2~120Hz. 2Hz~20 Hz in 1 Hz increment,
and 20~ 120 Hz in 5 Hz increment.
Press SET button to enter the Pulse Rate setting mode, then press
INCREASE or DECREASE button to adjust Pulse Rate to the setting
recommended by your prescrinbing medical professional.
5.11 Adjust the Pulse Width:
The pulse width is adjustable 50~300µs in 10µs increment.
Press SET button to enter the Pulse Width setting mode, then press INCREASE
or DECREASE button to adjust Pulse Width to the setting recommended by your prescribing medical professional.
5.12 Adjust Timer:
The timer is adjustable 5~90 minutes in 5 minute increments. Continuous option is just the next step to
90 minutes (i.e. from 5-90 minutes to continuous and then to 5 minutes is a cycle). During 5 minute final
count down, the clock symbol will flash once every one second.
5.6 Select the mode:
Press "MODE" button to set the stimulation mode recommended by
your physician or therapist. For details about stimulation modes, please
refer to Sec. 3 "Technical Specifications".
Caution: Consult physicians for your suitable stimulation mode.
5.7 Adjust the Ramp Time: (except the Constant Mode)
From 1 second to 8 seconds, the stimulator special circuitry is designed so the electrical impulses gradually build to a
peak and then level off. This "Ramped" pulse produces a gradual muscle contraction emulating natural muscle
movement. It can also prevent spastic patients from reacting adversely.
Press SET button to enter the Ramp setting mode, then press INCREASE or DECREASE button to adjust Ramp
Time to the setting recommended by your medical professional.
5.8 Adjust Contraction (ON) Time:(except the Constant Mode)
From 2 seconds to 99 seconds, press SET button to enter the ON Time setting mode, then press INCREASE or
DECREASE button to adjust On Time to the setting recommended by your prescribing medical professional.
1413