5.15 Patient Compliance Timer:
The patient compliance timer can memorize 60 sets of operation records; the
total record time is 999 hours.
After the unit is turned off, you can start to use patient compliance timer. First,
press and hold "Mode" button and turn on either amplitude knob
simultaneously to initiate patient compliance timer.
Individual treatment time:
Press "INCREASE" button (triangle button) or "DECREASE" button (inverted
triangle button) to see next record of treatment time with the number of times
or previous record of treatment time with the number of times.
Press and hold "Set" button for 3 seconds to delete the showing record. After
the record is deleted, the unit wil make a"beeping"sound.
NOTE:
1. If the treatment time is under one minute, it will not be recorded. For example, If your treatment time is 10 minutes and
30 seconds, the patient compliance timer will record 10 minutes, not 11 minutes.
2. The patient compliance timer can only record up to 999 minutes for each treatment. Therefore, if you keep using the
stimulator for over 999 minutes, it will only record 999 minutes and the record time will flash to mean the treatment time
is over 999 minutes.
5.13 Adjust Channel Amplitude:
Turn Channel 1 or 2 clockwise. The output indication will be showed on the left side (Channel 1) and right side
(Channel 2) of the LCD screen as long as the unit is in operation. Slowly turn the Channel Amplitude control until
you reach the setting recommended by your medical professional. Repeat for the other channel, if both channels are
to be used.
Caution: If the stimulation levels are uncomfortable or become uncomfortable, reduce the stimulation intensity to a
comfortable level and contact your physician if problems persist.
5.14 Turn Unit Off:
Turn both Channel Amplitude controls to "OFF". Then unplug the electrode lead wires, grasping them by the plug,
not the cord. If treatment will be resumed shortly, the electrodes may be left on the skin. When the electrodes are
removed, clean the skin thoroughly with mild soap and water. If there is skin irritation, consult your medical
professional.
Caution: When therapy time is completed, and if the user doesn't turn off the amplitude knob, the unit will continue the
"beeping" sound every 10 seconds until the amplitude knob is turned off completely.