AC40 Service Manual Page 48
Select “Transducer calibration”
Place the transducer on the 2cc 60318-5 (IEC126) coupler (Figure 19)
Figure 19
Adjust the “Adjustment” slider until the calibration level is obtained on the measurement amplifier
(i.e. 90 dB SPL) (Figure 20)
Figure 20
Note: Remember to add or subtract the correction for your microphone
Check the frequency accuracy with a frequency counter. The accuracy has to be better than ±1%