AC40 Service Manual Page 49
Press “Next”, when the calibration level is obtained, to get the next frequency
Repeat “Next” until all stimuli is calibrated, for both left and right transducer.
Write the calibration to Hardware
Select the Main menu (Figure 21)
Select “Write to hardware”
A dialog box will appear with the message “Calibration data successfully written to hardware”
Figure 21