Air Fx Irrigator Instructions for Use - US Page 25
Appendix A
Portable and mobile RF communications equipment can affect the Air Fx. Install and operate the Air Fx according to the EMC
information presented in this chapter.
The Air Fx has been tested for EMC emissions and immunity as a standalone instrument. Do not use the Air Fx adjacent to or
stacked with other electronic equipment. If adjacent or stacked use is necessary, the user should verify normal operation in the
configuration.
The use of accessories, transducers and cables other than those specified, with the exception of servicing parts sold by
Interacoustics as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the
device.
Anyone connecting additional equipment is responsible for making sure the system complies with the IEC 60601-1-2 standard.
Guidance and manufacturer’s declaration - electromagnetic emissions
The Air Fx is intended for use in the electromagnetic environment specified below. The customer or the user of the Air
Fx should assure that it is used in such an environment.
Electromagnetic environment - guidance
CISPR 11
The Air Fx uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The Air Fx is suitable for use in all commercial, industrial,
business, and residential environments.
Harmonic emissions
IEC 61000-3-2
Complies
Class A Category
flicker emissions
Recommended separation distances between portable and mobile RF communications equipment and the Air Fx.
The Air Fx is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the Air Fx can help prevent electromagnetic interferences by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Air Fx as recommended below, according to
the maximum output power of the communications equipment.
power of transmitter
[W]
Separation distance according to frequency of transmitter
[m]
d = 1.17
√
d = 1.17
√
d = 2.23
√
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHZ, the higher frequency range applies.
Note 2 These guidelines may not apply to all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Air Fx is intended for use in the electromagnetic environment specified below. The customer or the user of the
Air Fx should assure that it is used in such an environment.
level
Environment-Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
+8 kV air
+8 kV air
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be greater
Electrical fast transient/burst
IEC61000-4-4
+2 kV for power supply lines
+1 kV for input/output lines
+2 kV for power supply lines
+1 kV for input/output lines
Mains power quality should
be that of a typical
commercial or residential
environment.
IEC 61000-4-5
+2 kV common mode
+2 kV common mode
Mains power quality should
be that of a typical
commercial or residential