Guidance and manufacturer’s declaration - electromagnetic emissions
The Eclipse is intended for use in the electromagnetic environment specified below. The customer or the user of the Eclipse should
assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Eclipse uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Eclipse is suitable for use in all commercial, industrial, business,
and residential environments.
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Not applicable
Recommended separation distances between portable and mobile RF communications equipment and the Eclipse.
The Eclipse is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user
of the Eclipse can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Eclipse as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum output power
of transmitter
[W]
Separation distance according to frequency of transmitter
[m]
150 kHz to 80 MHz
d = 1.17
P
80 MHz to 800 MHz
d = 1.17
P
800 MHz to 2.5 GHz
d = 2.23
P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHZ, the higher frequency range applies.
Note 2 These guidelines may not apply to all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Eclipse is intended for use in the electromagnetic environment specified below. The customer or the user of the Eclipse should assure
that it is used in such an environment.
Immunity Test
IEC 60601 Test
level
Compliance
Electromagnetic
Environment-Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV contact
+8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be greater than 30%.
Electrical fast
transient/burst
IEC61000-4-4
+2 kV for power supply lines
+1 kV for input/output lines
Not applicable
+1 kV for input/output lines
Mains power quality should be that of a
typical commercial or residential
environment.
Surge
IEC 61000-4-5
+1 kV differential mode
+2 kV common mode
Not applicable
Mains power quality should be that of a
typical commercial or residential
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
lines
IEC 61000-4-11
< 5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
Not applicable
Mains power quality should be that of a
typical commercial or residential
environment. If the user of the Eclipse
requires continued operation during
power mains interruptions, it is
recommended that the Eclipse be
powered from an uninterruptable power
supply or its battery.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics Eclipse EP25 and is the answer not in the manual?

Interacoustics Eclipse EP25 Specifications

General IconGeneral
BrandInteracoustics
ModelEclipse EP25
CategoryMedical Equipment
LanguageEnglish

Related product manuals