Da Vinci Firefly Imaging System User Manual Addendum 552064-10 Rev. A
Table A.2 Firefly Sensitive Mode Troubleshooting (continued)
Symptom
Reason Action
Administration of flu-
orescent agent is
inadequate or is not
completed correctly.
Confirm that administration of the fluorescent
imaging agent is completed correctly.
Endoscope limitations. Use another Intuitive endoscope with Sensitive
mode capability.
A.2 Firefly Laser Test Procedure
If the Firefly laser is not operating, follow these steps to test it:
1. Ensure that the endoscope connector is fully inserted in the Endoscope Controller receptacle.
2. Hold the endoscope and point it at the test target on the Firefly Imaging Quick Reference
Guide.
3. Switch to Firefly imaging. If the laser is working and you are aiming the endoscope at the test
target, the target should appear lit green on the stereo viewer and touchscreen.