Service
1.Turnkeytoswitchmobilitydeviceoff.
2.
Pressandholdkeys.
3.Turnkeytoswitchmobilitydeviceon.
Mobilitydeviceenterssetupmodeaftertwoseconds.
Enternextdisplaywith
key.
Enternextdisplaywith
key.
Enternextdisplaywith
key.
Settingcounter
ThischapterappliesformobilitydeviceswithLCD
display.
1.EntersetupsettingsasdescribedinEnteringsetup(LCD
Display),page16.
Defaultsettingis0.Youcansetakmvalueas
serviceinterval.
Ifcountervalueisgreaterthansetvalue,
symbolwillashforoneminute,whenkeyisswitched
on.
+
0(OFF)+500+1000
+1500+2000..........
+9500+9999(max)
-
999995009000
85008000.........
5000(OFF)
2.
Pressandholdkeyfor2sec.toerasecounter.
3.
Presskeytoincreasevalue.
4.
Presskeytodecreasevalue.
5.
Presskeytosaveandenternextpage.
Settingbatterygauge
ThischapterappliesformobilitydeviceswithLCD
display.
Settingbatterygaugeismandatorywhenretrotting
operatingconsole.
Adjusttheparameterinbatterygaugetosuitdifferenttypes
andsizesofbatteries.
AGMbatteryGelbattery
(1)bigsize
batteries(>
50A)
(2)small
sizebatteries
(<50A,
default)
(3)smallsize
batteries(<
50A)
(4)bigsize
batteries(>
50A)
1.
Presskeytoincreasevalue.
2.
Presskeytodecreasevalue.
3.
Presskeytosaveandenternextpage.
Settingwheeldiameter
ThischapterappliestomobilitydeviceswithLCD
display.
Settingwheeldiameterismandatorywhen
retrottingoperatingconsole.
Setvalueaccordingtowheeldiameter11” ,12”,13”,14”or
15”forcorrectdisplayofspeed.