Questions and Answers

  • S
    samuel79Aug 17, 2025
    What to do if the motor is not running, but the mains indicator lights up and the relay clicks in my Invacare Medley Ergo?
    • C
      Cassandra RileyAug 17, 2025
      If the mains indicator lights up, but the motor of your Invacare Medical Equipment isn't running and you hear a clicking noise from the relay in the control unit, there are several potential causes. First, ensure the motor plug is fully inserted into the control unit. If it is not, insert the motor plug properly. Other possible causes include a defective motor, a damaged motor cable, or a defective control unit, in which case you should replace the motor, cable or control unit.
  • L
    Lisa LopezAug 20, 2025
    Why is the mains indicator not lighting up on my Invacare Medical Equipment?
    • J
      Jennifer SpencerAug 20, 2025
      If the mains indicator on your Invacare Medical Equipment does not light up, it could be due to a few reasons. First, check if the mains are properly connected and connect them if they are not. If the mains are connected, the fuse in the control unit may be blown, or the control unit itself may be defective. In these cases, replace the control unit.
  • S
    Samantha StoutAug 22, 2025
    How to troubleshoot a control unit that works in one direction, but not another on my Invacare Medley Ergo Medical Equipment?
    • M
      Michael KirbyAug 22, 2025
      If the control unit of your Invacare Medical Equipment is in order, except for one direction on one channel, this indicates a problem with either the control unit or the hand control. The solution is to replace the defective control unit or the defective hand control.
  • J
    James ChanAug 25, 2025
    What to do if the motor is running, but the piston rod does not move on my Invacare Medical Equipment?
    • G
      Gail WilsonAug 26, 2025
      If the motor is running in your Invacare Medical Equipment, but the piston rod isn't moving, it means the motor is damaged, and you should replace the motor.
  • E
    Emily JohnsonAug 27, 2025
    What to do if the motor of my Invacare Medley Ergo Medical Equipment is not running and the relay clicks?
    • G
      Gregory Brown MDAug 27, 2025
      If the mains indicator lights up, but the motor isn't running and you hear a clicking noise from the relay in the control unit of your Invacare Medical Equipment, it could be due to several reasons. First, ensure the motor plug is fully inserted into the control unit. If that's not the issue, the motor itself might be defective and require replacement. Another possibility is a damaged motor cable, which would also need replacing. Finally, the control unit could be defective, necessitating its replacement.
  • M
    Mark RobinsonAug 30, 2025
    What to do if the motor of my Invacare Medley Ergo is not running and there is no relay sound?
    • L
      lisa86Aug 30, 2025
      If the mains indicator on your Invacare Medical Equipment lights up, but the motor isn't running and there's no relay sound from the control unit, the issue likely lies within the control unit itself, meaning it may need to be replaced. Alternatively, the hand control could be defective and require replacement.
  • M
    Ms. Barbara CainSep 1, 2025
    How to troubleshoot the mains indicator not lighting up on Invacare Medical Equipment?
    • G
      gmezaSep 1, 2025
      If the mains indicator on your Invacare Medical Equipment does not light up, there could be a few reasons. First, check if the mains are properly connected and connect them if they aren't. Second, the fuse in the control unit might be blown, in which case you should replace the control unit. Finally, the control unit itself may be defective and require replacement.
  • M
    Matthew BeltranSep 4, 2025
    Why is one direction on one channel not working on my Invacare Medical Equipment?
    • S
      simpsonmarySep 4, 2025
      If the control unit of your Invacare Medical Equipment is in order except for one direction on one channel, the control unit could be defective and require replacement. Another possibility is that the hand control is defective and should be replaced.
  • N
    Nathan EnglishSep 7, 2025
    What to do if the motor runs but the piston rod doesn't move on Invacare Medley Ergo Medical Equipment?
    • T
      thomas20Sep 7, 2025
      If the motor is running on your Invacare Medical Equipment, but the piston rod isn't moving, the motor is likely damaged and needs to be replaced.
  • A
    Andrew ToddSep 9, 2025
    What to do if my Invacare Medley Ergo Medical Equipment bed has no electric function?
    • K
      Kevin FlynnSep 9, 2025
      If your Invacare Medical Equipment bed has no electric function, here are a few things to check: * Ensure the bed is properly plugged into the mains. * Verify the hand control is correctly connected to the control unit. * If these steps don't resolve the issue, the hand control, control unit, or cables may be defective, and you should contact your dealer or a qualified technician for assistance.

Summary

General Information

Safety and Warnings

Highlights critical safety precautions and warnings for operating the Medley Ergo bed.

Receiving the Bed

Assembly of the Bed

Provides step-by-step instructions for assembling the Medley Ergo bed and its components.

Wiring

Explains the correct procedure for connecting the electrical components and motors of the bed.

Operating the Bed

Operating the Hand Control Unit

Details how to use the hand control unit for adjusting the bed's various functions and positions.

Operating the Braking Castors

Explains the procedure for engaging and releasing the bed's braking castors for stability.

Emergency Lowering of Backrest and Thigh Section

Provides instructions for emergency lowering of specific bed sections in case of power failure.

Cleaning and Maintenance

Electrical System

Troubleshooting Electrical System

Provides a guide to diagnosing and resolving common issues related to the bed's electrical components.

Invacare Medley Ergo Specifications

General IconGeneral
Mattress Platform Width90 cm
Frame MaterialSteel
Adjustable BackrestYes
FootrestYes
Seat MaterialNylon
Backrest MaterialNylon