Service Manual Invacare
®
Meteor
25
7.2.2 Operating console and keypad
Disassembly
1. Remove plastic caps.
2. Loosen Allen screws and washers and remove.
3. Carefully lift up operating console.
4. Remove the caps from the turret, and dismantle the
turret.
5. Unscrew the operating circuit board.
6. Withdraw the conductor foil with plug.
7. Raise the double-layer keypad with a flat tool at any
one location, and pull off.
8. Remove any adhesive remains from the self-
adhesive foil.
Reassembly
1. Reassembly takes place in reverse order to disassembly.