1.8.1 Error codes and diagnostic codes
The drive electronics are capable of rectifying some errors automatically. In this case the status
display will cease to flash. Please switch the remote on and off several times. Wait approx. 5
seconds each time before switching the remote on again. If this does not rectify the error, locate
the error using the flash codes shown below.
FLASH
CODE
FAULT IMMEDIATE MEASURE FURTHER HELP
1
Module defective. -
Consult dealer.
2
Lifter raised or
lowered too far (seat
not at driving height)
If lifter is raised, lower in
stages until the status
display stops flashing. If
lowered too far, raise lifter
in stages until the status
display stops flashing. If at
all possible, only drive
when the seat is at driving
height.
-
Accessory error. -
Consult dealer.
3
Error on right motor.
Connection
loose/faulty.
Check plug-in connectors. Consult dealer.
4
Error on left motor.
Connection
loose/faulty or motor
faulty.
Check plug-in connectors. Consult dealer.