12
Getting Started, Continued
Updating the
Firmware
To update your Qubit
®
2.0 Fluorometer with the latest firmware
available from Invitrogen, follow these steps:
1. Download the latest firmware to your USB drive from our
website (www.invitrogen.com/qubit
).
2. Press the Version button located on the top portion of the
touch screen. The Version Screen is displayed.
3. Insert the USB drive into the USB port and press Update.
Note: A green dot on the Update button indicates that the instrument
recognizes the USB drive; a red dot indicates that the USB drive is not
inserted into the USB port or that the instrument does not recognize
the USB drive.
After the successful update of the firmware, the following
screen is displayed before the instrument returns to the Home
Screen. You can now proceed to setting your standards and
quantitating your samples.