To obtain the best results, follow the recommendations below.
For more information, see Critical Qubit
Assay Considerations
in the Appendix, pages 33–35.
Do not operate
the instrument in direct sunlight.
Wear gloves during sample handling.
Use the Qubit
®
2.0 Fluorometer at room temperature only
(22–28ºC).
Store all kit reagents at room temperature and insert all
assay tubes into the Qubit
®
2.0 Fluorometer only for as much
time as it takes for the instrument to measure the
fluorescence.
Do not hold the assay tubes in your hand before performing
a measurement.
Make sure that you have calibrated the Qubit
®
2.0
Fluorometer using the appropriate standards.
Incubate the tubes for the Qubit
DNA and RNA assays for
2 minutes after mixing the sample or standard with the
working solution.
Incubate the tubes for the Qubit
protein assays for 15
minutes after mixing the sample or standard with the
working solution.
If you are performing multiple readings of a single tube,
remove the tube from the instrument and let it equilibrate to
room temperature for 30 seconds before taking another
reading.
Visit www.invitrogen.com/qubit for additional application
notes, technical notes, citations, firmware updates, and a list
of validated Qubit
assays that have been tested using the
Qubit
®
2.0 Fluorometer.
Assay Tubes
for the
Qubit
®
2.0
Fluorometer
Only thin-wall, clear 0.5 mL PCR tubes are appropriate for use in
the Qubit
®
2.0 Fluorometer. Acceptable tubes include Qubit
®
assay
tubes (Cat. no. Q32856, 500 tubes) or Axygen PCR-05-C tubes
(VWR, part number 10011-830). The minimum assay volume must
be 200 L for an accurate read.