33
Appendix
Critical Qubit
Assay Considerations
How the
Qubit
®
2.0
Fluorometer
Calculates
Concentration
The Qubit
®
2.0 Fluorometer generates concentration data based on
the relationship between the two Standards used in calibration
(three for the Qubit
protein assay). The plot below shows the line
corresponding to the curve-fitting algorithm (a modified Hill plot)
used in the calculation of concentration data for the Qubit
RNA
assay. For reference, the positions of the Standards and a set of data
points from an actual experiment are shown superimposed onto the
line. This plot demonstrates that the curve-fitting algorithm gives
accurate values for quantitation.
Figure 1. The curve-fitting algorithm used to determine concentration in the
Qubit
RNA assay. Data for the Qubit
dsDNA BR, Qubit
dsDNA HS,
Qubit
ssDNA, Qubit
RNA BR, and Qubit
protein assays are generated
by similar algorithms.
Continued on next page