Troubleshooting, Continued
High Reading
Your sample is out of range. Use a sample that is less
concentrated.
View the Check Standards Screen to confirm that the values
for the samples fall between the values of the standards (see
page 18).
Ensure that the lid is closed while reading standards and
samples.
Prepare samples and standards according to the instructions
in the Qubit
assay kit you are using.
Ensure that the assay is performed entirely at room
temperature.