Calibrating the Qubit
®
2.0 Fluorometer, Continued
Checking
Standards
The Check Standards Screen displays the raw fluorescence values
for Standard #1, Standard #2, Standard #3 (if applicable), and the
last measurement taken for a given assay. These values can assist
you in making a judgment regarding the performance of the
Qubit
®
2.0 Fluorometer and whether or not the samples you used
were within range.
For the Qubit
dsDNA BR, Qubit
dsDNA HS, Qubit
ssDNA, Qubit
RNA, and Qubit
RNA BR assays, the reading
given by Standard #2 should be much higher than that of
Standard #1, and the reading of your sample should fall
somewhere between that of the two Standards.
For the Qubit
protein assay, the reading given by Standard
#3 should be much higher than that of Standard #2, and the
reading of your sample should be no more than 50% greater
than Standard #3.
To access the Check Standards Screen, press Check Stds. on
the Standards Screen. The raw fluorescence values for the
standards and the last sample will be displayed on the screen.