Questions and Answers

  • M
    Michele CookAug 19, 2025
    What to do if IPG Photonics Measuring Instruments show High Back Reflection error?
    • J
      Jose TaylorAug 20, 2025
      If the Measuring Instruments by IPG Photonics display a 'High Back Reflection' error, it means the back reflected power has exceeded the pre-installed maximal applicable level. First, check that the focus position is correctly adjusted and that there are no surfaces that could cause high back reflection. Then, reset the error using the RS-232 command (“RERR”) or the touch-screen display on the front panel.
  • R
    Richard RyanAug 21, 2025
    Why do IPG Photonics Measuring Instruments show Overheat error?
    • C
      Christopher AndersonAug 21, 2025
      If you see a 'Overheat' error on your IPG Photonics Measuring Instruments, it indicates that the case temperature of the laser is too high. Check if the outside conditions are within the specified range and ensure sufficient airflow. The error message should disappear once the case temperature drops back within the operating range.
  • D
    Dawn DavisAug 25, 2025
    What to do if IPG Photonics Measuring Instruments show Power Supply Failure 2 error?
    • A
      Ashley ZunigaAug 25, 2025
      If your IPG Photonics Measuring Instruments display a 'Power Supply Failure 2' error, it means that while the internal main power supply is on and voltage is applied to the laser module, the voltage value is outside the expected range. Try resetting the error and ensure the internal main power supply is switched on. If the error continues, contact IPG Photonics for assistance.
  • T
    Taylor LongAug 26, 2025
    What to do if IPG Photonics Measuring Instruments show Power Supply Failure error?
    • B
      Benjamin BeckerAug 26, 2025
      If your IPG Photonics Measuring Instruments displays a 'Power Supply Failure' error, it indicates that, despite the internal main power supply being switched on, no voltage is being applied to the laser module. Try resetting the error. If the error persists, contact IPG Photonics for assistance.
  • J
    jaybeckAug 28, 2025
    What to do if IPG Photonics Measuring Instruments show Module Disconnected error?
    • J
      Jacqueline OliverAug 28, 2025
      If your IPG Photonics Measuring Instruments show a 'Module Disconnected' error, it means that digital data communication with the laser module inside the device is broken. Try resetting the error. If the error appears again, contact IPG Photonics for assistance.
  • J
    Jon FoxAug 30, 2025
    What to do if IPG Photonics Measuring Instruments show Duty Cycle Too High error?
    • J
      jonathanhughesAug 30, 2025
      If your IPG Photonics Measuring Instruments shows 'Duty Cycle Too High' error, the duty cycle is the percentage of how long the laser is in the “on” state in the given modulated period. Check the modulation signal and try to reset the error. If it appears again, contact IPG Photonics for assistance.
  • A
    Anita HughesSep 1, 2025
    What to do if IPG Photonics Measuring Instruments show Internal Communication Error?
    • G
      gutierrezmichaelSep 1, 2025
      If your IPG Photonics Measuring Instruments display a 'Internal Communication Error', it occurs when there is no connection to the Controller Area Network (CAN). Try to reset the error. If it appears again, contact IPG Photonics for assistance.
  • K
    klyonsSep 3, 2025
    What to do if IPG Photonics Measuring Instruments show High Pulse Energy error?
    • D
      Diane BarronSep 3, 2025
      If your IPG Photonics Measuring Instruments show a 'High Pulse Energy' error, check the modulation signal. Try to reset the error. If it appears again, contact IPG Photonics for assistance.
  • M
    Matthew JenningsSep 5, 2025
    What to do if IPG Photonics Measuring Instruments show Pulse Too Long error?
    • J
      James LewisSep 6, 2025
      If your IPG Photonics Measuring Instruments display a 'Pulse Too Long' error, check the modulation signal. Try to reset the error. If it appears again, contact IPG Photonics for assistance.
  • J
    Jimmy EdwardsSep 7, 2025
    What to do if IPG Photonics Measuring Instruments show Average Power Too High error?
    • M
      Molly SchneiderSep 7, 2025
      If your IPG Photonics Measuring Instruments display a 'Average Power Too High' error, check the modulation signal. Try to reset the error. If it appears again, contact IPG Photonics for assistance.

Summary

Overview of the YLR-Series Fiber Lasers

Safety Information and Conventions

Explains safety symbols and conventions used in the guide.

General Safety Instructions

Provides essential safety recommendations for operating the laser.

Using Your Device

Operation Control Modes

Explains the differences between Local and Remote control modes for the laser.

Computer Interface;Commands

Interface Commands

Lists and explains the ASCII commands used for controlling the laser via interface.

Pulse Shaping

Using the Pulse Shaper Program

Explains how to use the Pulse Shaper interface and its various tabs.

Shape Editor Tab

Details on creating and editing pulse shapes using the Shape Editor.

Sequence Editor

Instructions for creating and managing pulse sequences for the laser.

Troubleshooting

Error Messages on the Display and Status Bits

Lists common error messages and their corresponding solutions.

Web User Utility

Service

Optical Fiber Connector Inspection and Cleaning

IPG Photonics YLR-100 Specifications

General IconGeneral
TypeFiber Laser
Output Power100 W
Wavelength1070 nm
CoolingWater-Cooled
Average Power100 W
Beam QualityM² < 1.1
Mode of OperationCW