Doc. SIE10166 Rev. 10 Page 23 of 28
. N. 1 Connection cable (6), 10 m long.
. N. 1 Rotary transducer (7) (or linear transducer, or both).
The rotating transducer comes with its flexible joint. The sketch of the joint is the following one.
3.13 HALL EFFECT CLIP-ON TRANSFORMER, CODE PII29166.
Hall effect clip-on transformer: it allows measuring the DC current of motors and of the auxiliary
supply. Main characteristics:
. Metering: AC and DC currents.
. DC measurement null with a knob.
. Ranges: 10 mV/A, 80 A DC, 40 A AC maximum, and 1 V/A, 2 A DC, 1.5 A AC maximum.
. Low battery indicator.
. Measurement errors: 4% of reading + 20 mA for the 80 A range; 2% of reading + 5 mA for the 2
A range.
. Phase shift (up to 65 Hz): maximum 1°.
. Maximum working voltage: 600 V rms.
. Power supply: alkaline 9 V battery, type 6 LR 61.
. Service life: 70 h typical.
. Maximum cable diameter: 10 mm.
. Weight: 330 g.
. Dimensions: 65 mm wide (clamp closed); 36 mm thick; 230 mm long.