Relative command STATus: OPERation:ENABle
Bit definition of operation event register:
Bit position 7 6 5 4 3 2 1 0
Bit name no use no use no use CV CC WTG no use CAL
Bit weight 16 8 4 1
STATus:OPERation:CONDition?
This command can read the parameter from the operation condition register. When the
parameter of operation condition register changes, the bit corresponding in operation event
register is 1.
Query syntaxSTATus: OPERation: CONDition?
Parameter:None
Returned parameter<NR1>
STATus:OPERation:ENABle
This command can set the parameter of operation event enable register. Setting parameter
can determine which bit value of operation event register is 1 and the bit will cause OPER of
status byte register to be 1.
Command syntaxSTATus: OPERation:ENABle <NR1>
Parameter:0~255
ExampleSTATus: OPERation:ENABle 128
Query syntaxSTATus: OPERation:ENABle
Returned parameter<NR1>
STATus:OPERation:NTRansition
This command is used to edit the negative transition trigger register of operation event. The
parameter determines which bits in operation event register is 1 will cause the OPER bit of
status byte register to be set to 1.
Command syntaxSTATus:OPERation:NTRansition <NR1>
Parameter:0~255
ExampleSTATus:OPERation:NTRansition 128
Query syntaxSTATus:OPERation:NTRansition