parameter0|1|ON|OFF, ‘6512’
exampleCAL:SEC 0, ‘6512’; CAL:SEC 1
inquiry syntaxCALibration:SECure
parameterNone
CALibration:VOLTage:LEVel
This command is used to define the voltage calibration point.P1 and P2 calibration points
should be calibrated according to priority.
Command syntaxCALibration:VOLTage:LEVel <point>
parameterP1|P2
CALibration:VOLTage [:DATA] {<numeric value>}
This command is used to read the actual voltage of the present calibration point.
Command syntaxCALibration:VOLTage [:DATA] <NRf>
parameter<NRf>
exampleCAL:VOLT 80.0002V
CALibration:CURRent:LEVel
This command is used to define the current calibration points. P1 and P2 calibration points
should be calibrated according to priority.
Command syntaxCALibration:CURRent:LEVel <point>
parameterP1|P2
CALibration:CURRent [:DATA] {<numeric value>}
This command is used to read the actual current of the present calibration point.
Command syntaxCALibration:CURRent [:DATA] <NRf>
parameter<NRf>
exampleCAL:CURR 3.002A
CALibration:SAVe