Calibration Command
Copyright© Itech Electronics Co., Ltd. 98
Example
CAL:VOLT 12.0V
CALibrate:CURRent:LEVel
This command is used to define the current calibration points.
Command syntax
CALibrate:CURRent:LEVel <P0|P1>
Arguments
P0|P1
Example
CAL:CURR:LEV P0
CALibrate:CURRent[:DATA]
This command is used to read the actual current of the present
calibration point.
Command syntax
CALibrate:CURRent[:DATA] <NRf>
Arguments
NRf
Unit
A
Example
CAL:CURR 20.0
CALibrate:MONitor:VOLTage:LEVel
This command is used to set monitoring the voltage calibration point.
Command syntax
CALibrate:MONitor:VOLTage:LEVel <P0|P1>
Arguments
P0|P1
Example
CAL:MON:VOLT:LEV P0
CALibrate:MONitor:VOLTage[:DATA]
This command is used to set the voltage calibration point data for monitoring.