Output Commands
Copyright© Itech Electronics Co., Ltd. 35
[SOURce:]VOLTage:RISE[:LEVel]
This command is used to set rising time of voltage.
Command syntax
[SOURce:]VOLTage:RISE[:LEVel] <MINimum|MAXimum|NRf>
Arguments
MINimum|MAXimum|NRf
Unit
S
Example
VOLT:RISE 5.0
Query syntax
[SOURce:]VOLTage:RISE[:LEVel]? [MINimum|MAXimum]
Returns
NRf
[SOURce:]VOLTage:FALL[:LEVel]
This command is used to set falling time of voltage.
Command syntax
[SOURce:]VOLTage:FALL[:LEVel] <MINimum|MAXimum|NRf>
Arguments
MINimum|MAXimum|NRf
Unit
S
Reset Value
0.001
Example
VOLT:FALL 2.0
Query syntax
[SOURce:]VOLTage:FALL[:LEVel]? [MINimum|MAXimum]
Returns
NRf