EasyManua.ls Logo

ITech IT6522D

ITech IT6522D
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Output Commands
Copyright© Itech Electronics Co., Ltd. 39
[SOURce:]POWer:RISE[:LEVel]
This command is used to set rising time of power.
Command syntax
[SOURce:]POWer:RISE[:LEVel] <MINimum|MAXimum|NRf>
Arguments
MINimum|MAXimum|NRf
Unit
S
Example
POW:RISE 5.0
Query syntax
[SOURce:]POWer:RISE[:LEVel]? [MINimum|MAXimum]
Returns
NRf
[SOURce:]POWer:FALL[:LEVel]
This command is used to set falling time of power.
Command syntax
[SOURce:]POWer:FALL[:LEVel] <MINimum|MAXimum|NRf>
Arguments
MINimum|MAXimum|NRf
Unit
S
Reset Value
0.001
Example
POW:FALL 3.0
Query syntax
[SOURce:]POWer:FALL[:LEVel]? [MINimum|MAXimum]
Returns
NRf

Table of Contents

Other manuals for ITech IT6522D

Related product manuals