Acceptance and Installation 
  Copyright © Itech Electronic Co., Ltd.  10 
Introduction of Binding Posts 
The maximum rated current of the terminal at position (A) is the maximum rated 
input current of the instrument. Securely fasten all wires by hand-tightening the 
binding posts. You can also insert standard banana plugs into the front of the 
connectors as shown in (B), and the maximum rated current at (B) is 10 A. 
⚫  IT6722/IT6722A: 
 
 
⚫  IT6723/IT6723B/IT6723C/IT6724/IT6724B/IT6724C: 
 
 
Connecting the DUT 
DUT connection is given below taking local measurement as example. For 
details of local and remote measurements, refer to “Remote sense 
function”. 
1.  Before connecting the DUT, be sure that the instrument Power is in Off 
position. 
2.  Check whether the shorting clip of Sense terminal is correctly 
mounted. 
3.  Unscrew the screws of the output terminals and connect the red and 
black test lines to the output terminals. Re-tighten the screws. 
When maximum current that one test line can withstand fails to meet 
the current rated current, use several pieces of red and black test lines. 
For example, the maximum current is 1,200A, then 4 pieces of 360A 
red and black lines are required. 
4.  Directly connect the other end of the red and black lines to the DUT 
terminal.