Function and Features 
  Copyright ©ITECH Electronics Co., Ltd.  24 
the display of this value. 
Remote sense function 
Remote sense can adjusted at the output voltage of the device under test, this 
feature allows to compensate the voltage drop on the wire between the front 
panel terminals of the power supply and the device under test. 
  Use local sense: 
Local sense doesn’t compensate the voltage drop on the connection wire, the 
operation is: 
1.  Use  the  short  clips  on  the  back  panel  of  the  instrument,  or  install  wire 
between + and S+, - and S-. 
2.  Connect the positive and negative terminals of the front panel to the device 
under test. 
  Use remote sense: 
Remote sense compensate the voltage drop on the connection wire, the 
operation is: 
1.  Disconnect the wires/short clips between + and S+, - and S-. 
2.  Connect wires from S+, S- to the device under test. 
3.  Connect wires from +, - to the device under test. 
 
 
NOTE 
In order to ensure the stability of the system, using armored twisted pair cable between 
the remote sense terminal of IT6900 and load. 
Please note that the positive and negative polarity when wiring, otherwise it will damage 
the instrument! 
3.12.2 RS485 and Output Sync signal interface 
Terminals on IT6922B/IT6932B/IT6942B/IT6952B/IT6953B power supply rear 
board comprise Output Sync signal interface and RS485 communication 
interface, as shown below.