SOURce Subsystem
Copyright © Itech Electronic Co., Ltd. 85
*RST Value
MAXimum
MINimum
Examples
VOLT:HIGH 5
Query Syntax
[SOURce:]VOLTage:HIGH?
[SOURce:]VOLTage:LOW?
Returned Parameters
<NR3>
[SOURce:]VOLTage:TRIGgered
This command sets the voltage level that will become active when the next
trigger occurs.
Command Syntax
[SOURce:]VOLTage[:LEVel]:TRIGgered <NRf+>
Parameters
MINimum to MAXimum | MAXimum | MINimum | DEFault
Unit
V (volts)
Returned Parameters
MAXimum
Examples
VOLT:TRIG 80 VOLT:LEV:TRIG 50
Related Commands
VOLT
[SOURce:]RESistance
This command sets the resistance of the load when operating in constant
resistance mode.