LCD display renewal rate setting
EA-5J converts the analog signal from the sensor to digital signal every 20 ms. The
digital signal is calculated automatically except the maximum and the minimum value
according as following four kinds of renewal rate for the length of time.
()LCD display : Renewal rate number of points
()0.2 : 0.2 seconds 10 points
()1.0 : 1.0 second 50 points
()5.0 : 5.0 seconds 250 points
()10.0 : 10.0 seconds 500 points
()Default setting: 0.2
LCD display shows the calculated average value except for the maximum and minimum at
every selected time base.
LCD display
Measuring time setting
The measuring time for ion balance can be chosen from following four kinds of
setting.
LCD display : Measuring time
10 : 10 seconds
30 : 30 seconds
60 : 60 seconds
con : Continuous
()Default setting: 10
LCD display
-page16-