32
9. Product Specifications
9.1 Delivery forms
– Programat CS4
– Power cord
– Sintering Furnace Temperature Checking Set Starter Kit
– Operating Instructions
– Sinter tray
– Sinter tray fork
– IPS e.max Speed Crystallization Tray
– USB download cable
– var. accessories
9.2 Technical data
Power supply 118–240 V / 50–60 Hz
Overvoltage category II
Contamination level 2
Acceptable voltage fluctuations ± 10 %
Max. power consumption 16 A at 118 V
8 A at 240 V
Electric fuses 250 V / T16 A high breaking (heating circuit)
Dimensions of electric fuses 250 V / T16 A high breaking (heating circuit)
Dimensions of the closed furnace Depth: 430 mm
Width: 310 mm / 390 mm (with cooling tray)
Height: 570 mm
Usable size of the firing chamber Diameter: 80 mm
Height: 80 mm
Max. firing temperature 1600 °C/2912 °F
Weight Furnace base: 10 kg
Furnace head: 17 kg
Safety notes
The sinter furnace complies with the following standards:
– IEC 61010-1: 2010
– EN 61010-1: 2010
– UL 61010-1: 2012-2015
– CSA 61010-1: 2012 – 2015
– IEC 61010-2-010: 2014
– EN 61010-2-010: 2014
– UL 61010-2-010: 2015
– CSA 61010-2-010: 2015
Radio protection / electromagnetic compatibility EMC tested
9.3 Acceptable operating conditions
– Acceptable ambient temperature range: +5 °C to +40 °C (+41 °F to +104 °F)
– Acceptable humidity range: Relative humidity 80% for temperatures up to 31 °C/88 °F
gradually decreasing to 50% relative humidity at 40 °C/104 °F,
condensation excluded.
– Acceptable ambient pressure: The furnace is tested for use at altitudes of up to 2000 m
above sea level.
9.4 Acceptable transportation and storage conditions
– Acceptable temperature range: –20 °C to +65 °C (–4 °F to 149 °F)
– Acceptable humidity range: Max. 80% relative humidity
– Acceptable ambient pressure: 500 mbar to 1060 mbar
Use only the original packaging together with the corresponding foam material for shipping purposes.