JWI-700D/700S 11
always: Tare/Zero function acts by pressing down key TARE or ZERO even if it is not
stable
12. Power saving function
on: Enable the power saving function
off: Disable the power saving function
13. Initialization
Press M-/SET, then press M+/SAVE to start initialization. After the window displays “ok”,
the initialization is finished
6. Calibration
Note: Before calibration, please set the unit first. The unit used in calibration is the one that has
been set before. During the calibration procedure, press ZERO/ESC to return to normal
weighing mode without saving.
Here we take 3kg/10g as an example
1. Press and hold TARE while powering on. Do not release it till the window displays “CAL”.
2. With no load on the weighing pan, press TARE to start zero point calibration. “ ”is
blanking at the left bottom.
3. Wait till the window displays the first calibration value. “ ”appears at the left bottom.
Note: The first calibration value is default. If you need to change the value, do as the following:
Press M-/SET to enter the value setting. Press MR/ or TARE/ to move the key leftwards
or rightwards. Press +/PRINT/HI or -/HOLD/LO to change the value. Press M+/SAVE to
save.
4. Put the corresponding weight on the weighing pan, and then press TARE to complete the
first point calibration. “ ”appears at the left bottom.
Note: After the first point calibration, the window can display the weight value. If no need for
the other point calibration, move to step 6 to finish the calibration procedure.
5. Add another mass to the current weight. The window will show the total weight. Press TARE