EasyManua.ls Logo

Jasco V-550 - 9 Environment]; Hardware Setting]

Default Icon
99 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
72
9. [Environment]
S ystem hardw are setting, self-diagnosis, optional accessory setting, and w avelength correction can be
started from this m enu.
In the [S pectra M anager] w indow , double-click [E nvironm ent]. T he [E nvironm ent] dialog box show n below
appears. C lick an item .
F igure 9.1 [E nvironm ent] dialog box
[Item s] Lists the item s to set.
[H ardw are S etting]: S ets w hether to turn the lam p O N or O F F , selects the lam p changeover
w avelength, and sets the lam p operating hours
[D iagnosis]: D iagnoses the spectrophotom eter status.
[A ccessories S etting]: S ets the conditions of the optional intelligent accessories that have
com m unicating functions.
[W avelength
C orrection]: C orrects the w avelength using the em ission of the deuterium lam p.
9.1 [H ardw are S etting]
S ets w hether to turn the lam p O N or O FF, selects the lam p changeover w avelength, and sets the lam p operating
hours. Select [H ardw are S etting] from the [Environm ent] dialog box, then press <E xecute>. The dialog box show n
below appears.
F igure 9.2 [H ardw are S etting] dialog box
[Lam p] S ets w hether to turn the deuterium and halogen lam ps O N or O F F. S elect
a check box, then press <O K > to close the dialog box. T he selected lam p

Table of Contents