Instruction Manual
500 795/REV E/03-17
Flame Photometer
Models PFP7 and PFP7/C

Other manuals for jenway PFP7

Questions and Answers

  • V
    Victor BaldwinJul 31, 2025
    What to do if the flame will not light on jenway PFP7 Measuring Instruments?
    • L
      Lisa LawrenceJul 31, 2025
      If the flame won't light on your Jenway Measuring Instruments, start by filling the drain 'U' tube with deionized water. Check if the fuel cylinder is empty or if the regulator pressures are incorrectly set; replace the cylinder or reset the regulator as needed. Ensure an adequate air supply by checking the air connection and the compressor's operation. If the propane/butane fuel restrictor is blocked, carefully 'blow out' the restrictor, avoiding the use of wire to prevent damage.
  • Y
    Yvonne MckenzieAug 4, 2025
    Why am I getting unstable results with my jenway PFP7?
    • C
      Christy McdanielAug 4, 2025
      Unstable results with your Jenway Measuring Instruments can stem from several factors. Condensation in the air supply may be the culprit; try draining the tubing and consider using a water separator for a moisture-free air supply. The nebulizer could be blocked, so check its performance. Ensure the fuel valve is correctly set for the element being determined. A nearly empty fuel cylinder can also cause issues, so replace it with a full one. Relocate the instrument if it's in a strong draught or very bright sunlight. Also, ensure the 'U' tube is clear and the drain trap is correctly positioned. Atmospheric contamination can cause unstable results, so remove the source or improve ventilation. Ensure the instrument is always flushed with deionized water before shutdown.
  • S
    Shelby AcostaAug 8, 2025
    Why are my jenway PFP7 Measuring Instruments giving non-linear results?
    • S
      Steven JohnsonAug 8, 2025
      Non-linear results on your Jenway Measuring Instruments can arise from several causes. A blocked nebulizer might be the issue, so check its performance. Contaminated deionized water can also lead to this problem, so make up new standards in fresh water. Incorrect dilution or the use of an incorrect calibration standard can also cause non-linear results; prepare fresh dilutions of the calibration standard. Ensure the fuel valve is correctly set. Additionally, verify that the instrument is draining adequately by checking the 'U' tube for blockages and ensuring the drain trap is at the correct level.
  • D
    Diane ValentineAug 13, 2025
    How to fix the 'unable to set standard reading' error on Jenway Measuring Instruments?
    • C
      Chelsea ThompsonAug 13, 2025
      If you're unable to set a standard reading on your Jenway Measuring Instruments, the calibration standard might be at an incorrect concentration or incorrectly diluted. In this case, make new standards to check the calibration procedure. Alternatively, the blank calibration standard may be contaminated or incorrectly set; re-make the standard and re-set the blank.
  • I
    ibakerAug 17, 2025
    What to do if there is no reading on the display of Jenway Measuring Instruments?
    • G
      gene34Aug 17, 2025
      If there is no reading on the display of your Jenway Measuring Instruments, first ensure that the correct filter is selected. If not, select the correct filter. If the calibration standard was incorrectly made, remake the standard. A completely blocked nebulizer can also cause this issue, so check the nebulizer performance.
  • W
    wpaceAug 21, 2025
    Why is my jenway PFP7 Measuring Instruments getting no electrical power?
    • D
      daviskellyAug 21, 2025
      If you're experiencing a lack of electrical power with your Jenway Measuring Instruments, start by ensuring the AC supply is fully engaged by checking that the power lead is correctly inserted. If the issue persists, verify that an AC supply is available by checking the mains supply.
  • D
    Daniel VelasquezAug 26, 2025
    What causes the 'unable to set blank' error on jenway PFP7 Measuring Instruments?
    • C
      Caroline MayerAug 26, 2025
      If you're unable to set a blank on your Jenway Measuring Instruments, it could be due to a contaminated blank. Try remaking the blank and re-checking. Another possible cause is a burner encrusted with salt; clean the burner and ensure that the cleaning procedure is carried out more frequently.

Summary

Safety Information

Prevention of Fire and Electric Shock

Warnings regarding fire and electrical hazards, emphasizing authorized repair by qualified personnel.

General Safe Operating Practice

Guidelines for laboratory practices, chemical handling, and hazard assessment.

Introduction

Instrument Description

Overview of the PFP7 and PFP7/C models, their purpose, and features.

Specification

Technical details including ranges, limits of detection, reproducibility, and operating conditions.

Principles of Operation

Explanation of flame photometry and the device's basic components.

Good Practice Guidelines

Recommendations for optimal instrument performance and sample handling.

Installation

Services Required

Details on necessary fuel, air supply, voltage, and drain connections for setup.

Unpacking

List of components to check upon receiving the instrument and accessories.

Mixing Chamber Assembly

Step-by-step instructions for assembling the mixing chamber, burner, and nebuliser.

Installation

Procedures for connecting air and fuel, power, and ensuring proper ventilation.

Accessories

Information on optional parts like air compressors and water separators.

Analysis Preparation

Calibration Standards

Information on available calibration standards and their preparation.

Sample Preparation

Practical advice for preparing samples to ensure accurate analysis.

Dilution

Guidance on diluting samples and standards for optimal flame photometry readings.

Operation

Front and Rear Panel Controls

Description of all controls and connectors on the front and rear panels.

Flame Ignition and Optimisation

Steps for igniting the flame and optimizing performance.

Calibration: PFP7

Instructions for calibrating the PFP7 model using standard solutions.

Calibration: PFP7;C

Steps for calibrating the PFP7/C model for direct concentration readings.

Shutdown

Procedures for safely turning off the instrument after use.

Operating Precautions

Safety warnings and recommendations during instrument operation.

Accessories

Water Separator

Details on the water separator, its function, and specifications.

Compressor

Information on the air compressor, its specifications, and installation.

Maintenance

General

Overview of minimal maintenance requirements for sustained performance.

Weekly Maintenance

Procedure for cleaning the nebuliser to ensure efficient operation.

Monthly Maintenance

Routine maintenance tasks, including weekly schedule and cleaning solution use.

Six-monthly Maintenance

Procedure for cleaning the burner and associated components.

Trouble Shooting

General

Step-by-step guide for resolving common operational faults.

Unstable Results

Potential causes and solutions for inconsistent readings.

Unable to Set Standard Reading

Troubleshooting steps for calibration issues.

Non-linear Results

Causes and fixes for non-linear measurement outcomes.

Flame Will Not Light

Solutions for ignition failures.

No Electrical Power

Steps to take when the instrument is not receiving power.

Unable to Set Blank

Troubleshooting for issues with setting the blank reading.

No Reading on Display

Steps to take when the display shows no output.

Spares

Minor Spares Kit (PFP7;MINKIT)

Description and part codes for a kit of common replacement parts.

EU Declaration of Conformity

jenway PFP7 Specifications

General IconGeneral
Brandjenway
ModelPFP7
CategoryMeasuring Instruments
LanguageEnglish