Section 1
Introduction
1.1 Instrument description
The PFP7 and PFP7/C are low temperature, single channel emission flame
photometers
designed
for the routine determination of sodium (Na), potassium (K) and lithium (Li). Additional filters for
the determination of calcium (Ca) and barium (Ba) are supplied with the PFP7. Both versions
are
fitted with an automatic flame failure detection for user safety, making them ideal for use in
clinical,
industrial or educational applications. Model PFP7/C is specifically designed for use
in
clinical
applications. The in-built lineariser circuitry enables readings of both Na and K, at
normal
clinical
serum concentrations to be displayed directly in mmol/l. Serum samples must be
diluted
200:1 or
100:1 prior to presentation to the flame photometer.
1.2 Specification
120-160 mmol/L Na (linearized)
1% Coefficient of variation (C.V.) for 20 consecutive samples using
10ppm Na set to read at 50.0 (PFP7) or 140 (PFP7/C).
Better than 2% when concentration of 3ppm Na and K and 5ppm Li
are set to read 100.
Interference from Na, K and Li when equal in concentration to the
test element will be less than 0.5%
Better than 2% over 5 minutes when continuously aspirating 10ppm,
sample set to read 50.0. Zero drift better than 2% per hour after
warm
Between 2 and 6 ml/minute
Nominal 1.00 volt for readout of 100.0.
The flame must be alight for at least 15 minutes to ensure
achievement of the above stated specification.
90-125V or 190-250V @ 50/60Hz
Propane, butane, natural gas or
L.P.G.