Two fuses are fitted to the instrument. FS1 is the primary fuse provided
to
protect the instrument and any accessories connected to the
sockets. FS2 is to protect the electronic circuitry in the PFP7. Both
FS1 and FS2 should be of an anti-surge type.
A three pin receptacle for the AC mains supply.
Two three pin sockets for flame photometer accessories such as a
compressor. The power from these sockets is controlled by the
instrument power switch on the front panel.
Two 4mm sockets which will provide an analogue signal of
approximately 1 volt when the readout is 1000 digits or 100.0, 10.00,
1.000.
1/4" connector for fuel tubing.
IMPORTANT: Fuel supply must be regulated to 10-15 inches water
gauge (0.36
-0.54psi or 0.025-0.38kg/cm
2
. See section 2.1.2 for
information on different fuel types.
5/16" connector for air tubing.
IMPORTANT: Air pressure on the inlet to the instrument must not
exceed 30psi (2kg/cm
2
). See section 2.1.3.
4.2 Operation – Flame Ignition and Optimisation
1. Ensure that the drain trap is pushed fully down on its clip. Ensure that the drain trap
has
solution in it and that no air locks are present. If necessary, purge by adding
deionised
water and allowing the surplus to run
away.
2. Close the fuel valve by turning fully clockwise. To avoid damaging the valve, it
should
not
be
forced.
3. Turn the fuel valve the required number of turns anti-clockwise depending on the
fuel
being
used.
Fuel Number of
turns
Propane
3
Butane
4
Natural gas Fully
open
L.P.G.
3.5
4. Turn on the fuel supply at source i.e.
cylinder.
5. Switch on electrical power by depressing the power switch. If the air compressor is
powered separately, switch on the air compressor. Ensure that air is present
by
listening for the hissing created as it passes through the
nebuliser.