9 Error messages
Error 1 Pressure sensor failure or overpressure.
Error 2 PT100 probe failure or overtemp.
Error 3 External alarm. Pressure switch or safety thermostat trigered.
To reset the safety thermostat press its button number
23
The thermostat is adjusted in the factory it must not be ope-
rated except to reset it.
Error 4 Manual valve (16) opened during a cicle.
Error 5 Opened door during a cycle.
Error 6 Pressure over +0.05bar with the autoclave stopped.
Error 7 E2PROM failure.
Error 8 Out of limits temperature during the sterilization.
Error 9 Mains supply failure during cycle.
10 Spare parts
Presoclave II 80 litres. Part No. 4002136