Cleaning / repairing the unit
26
15
A
M
P
Circulator: Mains circuit breakers (resettable) 15 A
Cooling machine: Fuse T 10.0 A, dia.5 x 20 mm
The mains fuses (8b) on the rear of the unit may easily be
exchanged as shown on the left.
Warning:
Before exchanging the fuses, turn off the mains power switch and
disconnect the power plug from the mains socket!
Only use fine fuses with a nominal value as specified.
Example:
Manufacturer Supplier Type Order No.
Wickmann Wickmann G- fuse insert
T10,0A 5x20 mm
No. 19195
8. Cleaning / repairing the unit
Caution:
• Always turn off the unit and disconnect the mains cable from the
power source before cleaning the unit.
Prevent humidity from entering into the circulator.
• Electrical connections and any other work must be performed by
qualified personnel only.
To maintain the full cooling performance, clean the condenser
from time to time.
• Switch off the unit, disconnect mains power cable.
• Hold the venting grid, pull out and remove.
• Clean the ribbed condenser with a vacuum cleaner.
• Replace the venting grid.
• Switch on the unit.