Operating procedures
22
6. Operating procedures
6.1. Power connection
Caution:
Only connect the unit to a power socket with earthing contact (PE –
protective earth)!
The power supply plug serves as safe disconnecting device from the
line and must be always freely accessible.
Never operate equipment with damaged mains power cables.
Regularly check the mains power cables for material defects (e.g. for
cracks).
We disclaim all liability for damage caused by incorrect line voltages!
Check to make sure that the line voltage matches the
supply voltage specified on the identification plate.
Connect the circulator with mains power cable (10a) to
the mains outlet (17).
Connect the refrigerated circulator with mains power
cable (10b) to the mains socket.
6.2. Switching on / Start - Stop
Switching on:
Circulator and cooling machine may be turned on and off
with separate mains switches. The integrated control light
will illuminate to indicate that power has been applied.
The unit performs a self-test. All segments of the 4-digit
LED temperature DISPLAY and all indicator lights will
illuminate (as illustrated on the left).
The display "OFF" indicates the unit is ready to operate
(standby mode).
Start:
Press enter
for about 4 seconds.
The LED temperature DISPLAY indicates the actual bath
temperature.
Stop:
Press enter
for about 4 seconds.
Turn the unit off with the mains power switch.
L To save energy, turn off the cooling machine with the
mains switch (1b) whenever cooling is not required.