Menu functions
50
Menu level 1
Ð
Level 2 Parameter level
L Press the key if parameter is to be retained.
Correction function for parameters or values (prior
to OK).
Î
or
x The parameter flashes, switch by pressing
and
L On level 2 a (I) is indicated for internal or an (E)
for external.
Example:
Î
or
x The parameter flashes, switch by pressing
and
L >NO< Carry out an ATC calibration
L >YES< return to standard operation after
calibration.
Î
x The parameter flashes, switch by pressing
and
L A >1-point<, >2-point< or >3-point< calibration
can be carried out.
The selected calibration is indicated on level 2 by 1 or
2 or 3.
Î
The value >TMPVAL< is only indicated
L In addition the measured temperature value
>CALVAL X< is saved during the next step.
Î
x Integer digits flash, set by pressing
+
x Decimal digits flash, set by pressing
+
L If only a 1-point calibration is carried out, the following menu items are not
indicated anymore
Level 2 Parameter level
Î
The value is only indicated