Remote control
74
Command Parameter Response of circulator
IN_PAR_04 none CoSpeed - Band limit (max. difference between the
temperatures in the internal bath and external system).
IN_PAR_05 none Factor pk/ph0:
Ratio of max. cooling capacity versus max. heating capacity
IN_PAR_06 none Xp control parameter of the internal controller.
IN_PAR_07 none Tn control parameter of the internal controller.
IN_PAR_08 none Tv control parameter of the internal controller.
IN_PAR_09 none Xp control parameter of the cascade controller.
IN_PAR_10 none Proportional portion of the cascade controller.
IN_PAR_11 none Tn control parameter of the cascade controller.
IN_PAR_12 none Tv control parameter of the cascade controller.
IN_PAR_13 none Adjusted maximum internal temperature of the cascade
controller.
IN_PAR_14 none Adjusted minimum internal temperature of the cascade
controller.
IN_PAR_15 none Band limit (upper)
IN_PAR_16 none Band limit (lower)
IN_MODE_01 none Selected setpoint:
0 = SETP 1
1 = SETP 2
2 = SETP 3
3 = Last setpoint setting was carried out
through an external programmer
IN_MODE_02 none Selftuning type:
0 = Selftuning „off“
1 = Selftuning „once“
2 = Selftuning „alwayS“
IN_MODE_03 none Type of the external programmer input:
0 = Voltage 0 V to 10 V
1 = Current 0 mA to 20 mA
IN_MODE_04 none Internal/external temperature control:
0 = Temperature control with internal sensor.
1 = Temperature control with external Pt100 sensor.
IN_MODE_05 none Circulator in Stop/Start condition:
0 = Stop
1 = Start
IN_MODE_08 none Adjusted control dynamics
0 = aperiodic
1 = standard
IN_HIL_00 none Max. cooling power (%).
IN_HIL_01 none Max. heating power (%).