English 33
Appendix B
Specifications and Standards
DEXIS Titanium Sensor Specifications
SENSOR ARCHITECTURE Indirect converting dental IO X -ray sensor
1692 by 1324 pixels
19.5 μm pixel pitch
20+ visible lp/mm
X-RAY PARAMETERS Sensor can be used with dental X-ray generators in the range
of 60 to 70 kV; at minimal 40 μGy incident dose
SW ARCHITECTURE USB 2.0 compliant USB interface + application interface to
dental imaging software products
Operation system:
Microsoft Windows 7 with current service pack
Microsoft Windows 8.1with current service pack
Microsoft Windows 10 with current service pack
ELECTRICAL RATING DC 5V, 0.5W
CONNECTION TO PC USB 2.0 compliant / USB3.x compatible
PROTECTION AGAINST SHOCK Type BF
applied part
MODE OF OPERATION Continuous
METHOD OF STERILIZATION Sensor not suitable for sterilization
ENVIRONMENTAL CONDITIONS Humidity Air Pressure Ambient
Temperature
USAGE
DEXIS Titanium sensor is not suitable to be oper-
ated in oxygen rich and/or explosive environ-
ments
30% to
95%
700 to 1060
hPa
5° to 30°C
TRANSPORTATION AND STORAGE
Transport in the supplied protective package
10% to
95%
-40° to 70°C
EU CLASSIFICATION Class IIa medical device according to MDD
93/42/ECC
Protection against water/matter – IP 68