Instructions for use
INTRAsurg 300 / INTRAsurg 300 plus
Always be on the safe side.

Table of Contents

Questions and Answers

  • M
    Matthew SanchezAug 19, 2025
    What does error 29 mean on KaVo Medical Equipment?
    • B
      Brian GreerAug 19, 2025
      If the KaVo Medical Equipment displays 'error 29', it indicates that the motor hose is damaged. Check the motor hose and replace it if necessary.
  • T
    Trevor KempAug 21, 2025
    What to do if KaVo Medical Equipment displays warning 50 - motor is overloaded?
    • Q
      qowenAug 21, 2025
      If the KaVo Medical Equipment displays the message 'warning 50 - motor is overloaded', it means the motor is overloaded because the maximum permissible motor output was exceeded. The motor will be blocked for 10 minutes, so allow it to cool down.
  • S
    Stephanie MooreAug 23, 2025
    What to do if KaVo Medical Equipment displays remark 65 - motor is overloaded?
    • D
      danielle26Aug 23, 2025
      If the KaVo Medical Equipment displays the message '>>remark 65 -motor is overloaded', it indicates a motor overload warning. To resolve this, stop the treatment as soon as possible and allow the motor to cool off for a few minutes.
  • Z
    zachary31Aug 26, 2025
    How to troubleshoot KaVo Medical Equipment when nothing works?
    • B
      briandeckerAug 26, 2025
      If your KaVo Medical Equipment is not working, here are a few things to check: * Check and replace the fuse if it's blown. * Ensure the unit is switched on using the mains switch on the back. * Make sure both ends of the power line are securely plugged in. * Verify that the correct handpiece or angle piece is properly mounted and set on the device. * If the cause is unknown, try turning the machine off and on again.
  • E
    Erik BarnesAug 27, 2025
    What to do if KaVo Medical Equipment displays warning 36 - missing foot control?
    • S
      Steven HarrisAug 27, 2025
      If the KaVo Medical Equipment displays the message '>>warning 36 -missing foot control', it could be due to a defective footswitch or a poor connection, possibly from moisture. Try drying the footswitch and then turn the machine off and on.
  • B
    Brian PerkinsAug 29, 2025
    What to do if KaVo Medical Equipment displays remark 66 - motor not plugged?
    • E
      edgarevansAug 29, 2025
      If the KaVo Medical Equipment displays the message '>>remark 66 -motor not plugged', this indicates that the motor hose is not correctly inserted or the motor is not correctly connected to the motor hose. Ensure that the motor hose is properly inserted and the motor is correctly plugged in.
  • R
    Robert CoxAug 31, 2025
    What does remark 67 - motor not plugged mean on KaVo Medical Equipment?
    • M
      Mr. Timothy Coffey DDSAug 31, 2025
      If the KaVo Medical Equipment displays the message '>>remark 67 -motor not plugged', it means the disconnection of the motor from the hose was detected when the motor started. Plug the hose onto the motor and screw it tight. The display should disappear after eight seconds, after which the motor can be restarted.
  • A
    ashley31Sep 2, 2025
    Why does the KaVo Medical Equipment motor shut off during treatment?
    • F
      Frank JacksonSep 3, 2025
      If the KaVo Medical Equipment motor shuts off during treatment without an error message, it might be due to the wrong handpiece or angle piece being mounted or set on the device. Make sure the correct handpiece or angle piece is properly adjusted.
  • D
    Diane AndersonSep 4, 2025
    What does warning 49 mean on KaVo Medical Equipment?
    • J
      Jennifer RobbinsSep 4, 2025
      If you see 'Display: warning 49' on your KaVo Medical Equipment, it means the power consumption of the surgical motor was too high. Press the ENTER button to continue. You may also need to exchange the motor hose or the surgical motor.
  • O
    opalmerSep 6, 2025
    Why is the coolant insufficient in my KaVo Medical Equipment instrument?
    • S
      Sara StewartSep 7, 2025
      If the coolant in the instrument of your KaVo Medical Equipment is insufficient, it could be due to a few reasons: * The spray nozzles may be crusty or soiled, so clean them with the nozzle needle. * The pump hose might be sticky or crusty. * The glass bottle with saline solution may not be ventilated; ensure it is.

Summary

User Instructions

User Guide

Overview of user information and symbols.

Target Group

Defines the intended users of the document.

Service

Provides contact information for service and support.

Warranty Terms and Conditions

Outlines the terms and conditions of the product warranty.

Transportation and Storage

Instructions for safely transporting and storing the device.

Safety

Description of Safety Instructions

Explains warning symbols, structure, and danger levels.

Purpose - Proper Use

Details the intended use and operational guidelines.

Product Description

INTRAsurg 300;300 Plus Overview

Overview of the INTRAsurg 300/300 plus device components.

Controls

Description of the device's control panel and buttons.

Multifunctional Foot Control

Details the functions and operation of the foot control pedal.

Rating Plate for 300;300 Plus

Information presented on the device's rating plate.

Technical Data for the 300;300 Plus

Specifications and technical details of the device.

First Use

Electrical Connection

Instructions for connecting the device to the power supply.

Unpacking

Steps for safely unpacking the device and its components.

Adjust Mains Input Voltage

Procedure for setting the correct mains input voltage.

Mount Bottle Holder and Use Coolant Container

Guide for attaching the bottle holder and using coolant containers.

Connect Surgery Motor and Coolant Reservoir

Steps for connecting the motor and coolant reservoir.

Operation

Operation in General

General operating procedures and startup instructions.

Surgical Motors

Information and procedures for using surgical motors.

Free Use

Instructions for using the device in "Free Use" mode.

Program

Guide for using the device's programmable functions.

Functions Exclusively with 300 Plus Program Version

Special functions available only on the 300 Plus model.

Preparation Methods DIN EN ISO 17664

Cleaning

Procedures for cleaning the device components.

Disinfection

Instructions for disinfecting device parts.

Sterilisation in a Steriliser (DIN EN 13060)

Guidelines for sterilising the device using an autoclave.

Troubleshooting

Exchange the Pump Hose

Procedure for replacing the pump hose.

Exchange Coolant Hose

Instructions for replacing the coolant hose.

Device Settings Table

Table for recording device settings before and after service.

Accessories

Details on Electromagnetic Compatibility

Guidelines and Manufacturers Declaration - Electromagnetic Transmission

Declaration regarding electromagnetic emissions.

Guidelines and Manufacturers Declaration - Electromagnetic Resistance to Jamming

Declaration regarding electromagnetic interference immunity.

KaVo INTRAsurg 300 Specifications

General IconGeneral
BrandKaVo
ModelINTRAsurg 300
CategoryMedical Equipment
LanguageEnglish

Related product manuals