What does error 29 mean on KaVo Medical Equipment?
- BBrian GreerAug 19, 2025
If the KaVo Medical Equipment displays 'error 29', it indicates that the motor hose is damaged. Check the motor hose and replace it if necessary.
What does error 29 mean on KaVo Medical Equipment?
If the KaVo Medical Equipment displays 'error 29', it indicates that the motor hose is damaged. Check the motor hose and replace it if necessary.
What to do if KaVo Medical Equipment displays warning 50 - motor is overloaded?
If the KaVo Medical Equipment displays the message 'warning 50 - motor is overloaded', it means the motor is overloaded because the maximum permissible motor output was exceeded. The motor will be blocked for 10 minutes, so allow it to cool down.
What to do if KaVo Medical Equipment displays remark 65 - motor is overloaded?
If the KaVo Medical Equipment displays the message '>>remark 65 -motor is overloaded', it indicates a motor overload warning. To resolve this, stop the treatment as soon as possible and allow the motor to cool off for a few minutes.
How to troubleshoot KaVo Medical Equipment when nothing works?
If your KaVo Medical Equipment is not working, here are a few things to check: * Check and replace the fuse if it's blown. * Ensure the unit is switched on using the mains switch on the back. * Make sure both ends of the power line are securely plugged in. * Verify that the correct handpiece or angle piece is properly mounted and set on the device. * If the cause is unknown, try turning the machine off and on again.
What to do if KaVo Medical Equipment displays warning 36 - missing foot control?
If the KaVo Medical Equipment displays the message '>>warning 36 -missing foot control', it could be due to a defective footswitch or a poor connection, possibly from moisture. Try drying the footswitch and then turn the machine off and on.
What to do if KaVo Medical Equipment displays remark 66 - motor not plugged?
If the KaVo Medical Equipment displays the message '>>remark 66 -motor not plugged', this indicates that the motor hose is not correctly inserted or the motor is not correctly connected to the motor hose. Ensure that the motor hose is properly inserted and the motor is correctly plugged in.
What does remark 67 - motor not plugged mean on KaVo Medical Equipment?
If the KaVo Medical Equipment displays the message '>>remark 67 -motor not plugged', it means the disconnection of the motor from the hose was detected when the motor started. Plug the hose onto the motor and screw it tight. The display should disappear after eight seconds, after which the motor can be restarted.
Why does the KaVo Medical Equipment motor shut off during treatment?
If the KaVo Medical Equipment motor shuts off during treatment without an error message, it might be due to the wrong handpiece or angle piece being mounted or set on the device. Make sure the correct handpiece or angle piece is properly adjusted.
What does warning 49 mean on KaVo Medical Equipment?
If you see 'Display: warning 49' on your KaVo Medical Equipment, it means the power consumption of the surgical motor was too high. Press the ENTER button to continue. You may also need to exchange the motor hose or the surgical motor.
Why is the coolant insufficient in my KaVo Medical Equipment instrument?
If the coolant in the instrument of your KaVo Medical Equipment is insufficient, it could be due to a few reasons: * The spray nozzles may be crusty or soiled, so clean them with the nozzle needle. * The pump hose might be sticky or crusty. * The glass bottle with saline solution may not be ventilated; ensure it is.
Brand | KaVo |
---|---|
Model | INTRAsurg 300 |
Category | Medical Equipment |
Language | English |
Overview of user information and symbols.
Defines the intended users of the document.
Provides contact information for service and support.
Outlines the terms and conditions of the product warranty.
Instructions for safely transporting and storing the device.
Explains warning symbols, structure, and danger levels.
Details the intended use and operational guidelines.
Overview of the INTRAsurg 300/300 plus device components.
Description of the device's control panel and buttons.
Details the functions and operation of the foot control pedal.
Information presented on the device's rating plate.
Specifications and technical details of the device.
Instructions for connecting the device to the power supply.
Steps for safely unpacking the device and its components.
Procedure for setting the correct mains input voltage.
Guide for attaching the bottle holder and using coolant containers.
Steps for connecting the motor and coolant reservoir.
General operating procedures and startup instructions.
Information and procedures for using surgical motors.
Instructions for using the device in "Free Use" mode.
Guide for using the device's programmable functions.
Special functions available only on the 300 Plus model.
Procedures for cleaning the device components.
Instructions for disinfecting device parts.
Guidelines for sterilising the device using an autoclave.
Procedure for replacing the pump hose.
Instructions for replacing the coolant hose.
Table for recording device settings before and after service.
Declaration regarding electromagnetic emissions.
Declaration regarding electromagnetic interference immunity.