Performance Verification
7. Enter the voltm
eter reading into the table for checking DC voltage accuracy
without remote sense. (See Table 16 on page 17.)
8. Enter the volt
meter reading into the table for checking DC voltage readback
accuracy. (See Table 17 on page 18.)
9. Enter the CU
T Readback Voltage into the table for DC voltage readback
accuracy. (See Table 17 on page 18.)
10. Calculate t
he difference of the two measurements t aken in steps 8 and 9 and
enter the absolute value into the difference column of the table. (See Table 17
on page 18.)
11. Increase the CUT output voltage by 25% of the FS output voltage.
12. Repeat st
eps 7 through 11 until you complete testing a t 100% of FS.
This completes the check for one channel. If needed, return to step 1 to run
through
the check for the next channel.
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference 33